This instrument is our high-end volumetric gas adsorption instrument. To obtain useful information about micropore, it is important to measure adsorption isotherm from low relative pressure with accuracy.
The BELSORP-max is designed for wide range adsorption isotherm for surface area and pore size distribution analysis. It can measure adsorption isotherms from relative pressure as low as 1 x10-8 (N2 at 77K, Ar at 87K), using a 13.3Pa pressure transducer. Also AFSM, the new method for free space measurement, is applied and adsorption isotherm can be measured with high accuracy. Chemisorption option enables unattended chemisorption measurement from pretreatment with step programs.
| Measuring principle | Volumetric gas adsorption method+AFSM |
|---|---|
| Adsorptive | N2, Ar, Kr, NH3, CO2, H2, CO, O2, CH4 and other non-corrosive gases H2O, MeOH, EtOH, C6H6 and other non-corrosive vapor |
| Analysis port | STD mode (p/p0=10-4 - 0.997) 3 ports High accuracy mode (p/p0=10-4 - 0.997) 2 ports High accuracy mode (p/p0=10-8 - 0.997) 1 port (OP : 2 ports) |
| Specific surface area | 0.01m2/g and above (N2/77K) 0.0005 m2/g and above (Kr/77K) |
| Pore size distribution | 0.32 - 500 nm in pore diameter |
| Pressure transducer | 133 kPa (1000Torr) 1.33kPa (10Torr) 0.0133kPa (0.1Torr) |
| Thermostated air oven | 40oC |
| Dewar vessel | Volume 2.6L Holding time 60H |
| Sample cell | Approx. 1.8cm3 (Option: 5 cm3) |
| Vacuum pump | Turbo molecular drug pump + Fore vacuum pump(OP) Ultimate vacuum: 6.7 x 10-7 Pa and below |
| Vacuum gauge | Pirani + Cold cathode gauge (ATM - 5x10-7 Pa) |
| Physical | W565 x H850 x D580 mm, 84 kg (Not includes vacuum pump, computer and its peripheral) |
| Power | Single phase: AC100-120V or 200-240V / 1500VA (400VA is for vacuum pump) |